From: V Brazilian consensus guidelines for detection of anti-cell autoantibodies on hep-2 cells
1 - Harmonization to the international guidelines for the anti-cell autoantibodies on HEp-2 cells and adoption of the corresponding ICAP AC code. | |
2 - Creation of preliminary Brazilian AC codes (BAC) for the patterns recognized by the BCA and not addressed in the ICAP algorithm at this time. | |
3 - Maintenance of the original BCA conceptual arrangement of groups and subgroups of patterns in the classification tree. | |
4 - Adequacy of the test designation: anti-cell autoantibodies (ANA HEp-2) | |
5 - Attention to the reproducibility of different patterns, depending on the commercial HEp-2 cell substrate used. | |
6 - Alert that solid phase-based tests should not replace Indirect Immunofluorescence on HEp-2 cells as a screening test for anti-cell antibodies. | |
7 - Dilution of screening at 1/80. | |
8 -Titration of the conjugate for each new brand and lot of the same brand. | |
9 - Use of quality control strategies. | |
10 - Adequacy of the report to express the name of the pattern, ICAP code, titer, fluorescence pattern in each cellular compartment and classification of the chromosome metaphase plate |